CVE-2025-4395 affects Medtronic MyCareLink Patient Monitor models 24950 and 24952 and is rated CVSS 6.8 (Medium). The issue is a built-in user account with an empty password, which means a person with physical access to the monitor can log in without credentials and access or modify system functionality. CISA published the advisory on 2025-07-24 and later issued Update A on 2026-05-07. Medtronic characterized the findings as low-risk and said updates began rolling out in June 2025.
- Vendor
- Medtronic
- Product
- MyCareLink Patient Monitor model 24950
- CVSS
- MEDIUM 6.8
- CISA KEV
- Not listed in stored evidence
- Original CVE published
- 2025-07-24
- Original CVE updated
- 2026-05-07
- Advisory published
- 2025-07-24
- Advisory updated
- 2026-05-07
Who should care
Healthcare providers, patients using the affected MyCareLink Patient Monitor models, biomedical engineering teams, and Medtronic support/security contacts should pay attention. Security and operations teams that manage connected medical devices should confirm update status and follow vendor handling guidance.
Technical summary
The CSAF advisory describes a local, physical-access issue in Medtronic MyCareLink Patient Monitor model 24950 and model 24952. A built-in user account is configured with an empty password, allowing an attacker with physical access to authenticate with no password and interact with system functionality. The supplied advisory states the vulnerability requires physical tampering with the monitor, and Medtronic says the security update is delivered automatically when the monitor is connected to the internet.
Defensive priority
Medium
Recommended defensive actions
- Ensure the remote monitor is connected to the internet so it can receive Medtronic's automatic security update process.
- Maintain possession of the home monitor and prevent unauthorized physical access.
- Use only home monitors provided directly by a healthcare provider or a Medtronic representative.
- Continue prescribing and using monitors as intended, following Medtronic's guidance for the device.
- Contact Medtronic security at [email protected] if additional assistance is needed.
- Review Medtronic's security bulletin for the affected MyCareLink Patient Monitor models.
- Follow CISA guidance on securing Internet of Things devices and home network security.
Evidence notes
Source timing is based on the supplied CVE and CSAF metadata: initial publication on 2025-07-24 and Update A on 2026-05-07. The advisory text says the finding is low-risk, requires physical tampering, and that Medtronic began deploying security updates in June 2025. The CSAF lists affected products as MyCareLink Patient Monitor model 24950 and model 24952. No KEV listing was supplied for this CVE.
Official resources
-
CVE-2025-4395 CVE record
CVE.org
-
CVE-2025-4395 NVD detail
NVD
-
Source item URL
cisa_csaf
-
Source reference
Reference
-
Source reference
Reference
-
Source reference
Reference
-
Source reference
Reference
-
Source reference
Reference
-
Source reference
Reference
-
Source reference
Reference
Publicly disclosed in CISA ICS Medical Advisory ICSMA-25-205-01 on 2025-07-24, with Update A published on 2026-05-07. The advisory and linked vendor guidance describe the issue and update rollout status.