PatchSiren cyber security CVE debrief
CVE-2025-4394 Medtronic CVE debrief
CVE-2025-4394 describes an unencrypted filesystem on internal storage in Medtronic MyCareLink Patient Monitor devices. CISA says an attacker would need physical access to tamper with the monitor to read or modify files, and Medtronic reported the issue as a low-risk finding while deploying security updates beginning in June 2025.
- Vendor
- Medtronic
- Product
- MyCareLink Patient Monitor model 24950
- CVSS
- MEDIUM 6.8
- CISA KEV
- Not listed in stored evidence
- Original CVE published
- 2025-07-24
- Original CVE updated
- 2026-05-07
- Advisory published
- 2025-07-24
- Advisory updated
- 2026-05-07
Who should care
Hospitals, clinics, physicians, home-care teams, and patients who use or manage Medtronic MyCareLink Patient Monitor model 24950 or 24952 devices should care, especially where the monitor could be physically accessed outside a controlled environment.
Technical summary
The advisory describes cleartext storage on internal device storage: files on the monitor are not encrypted at rest, so someone with physical access could read and modify them. The supplied CVSS vector is AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H (6.8 Medium), which is consistent with a confidentiality, integrity, and availability impact if the device is tampered with. The issue aligns with CWE-312 (Cleartext Storage of Sensitive Information).
Defensive priority
Medium. The attack requires physical access, which narrows exposure, but the potential impact to device data integrity and confidentiality is high enough to justify timely verification of update status and physical safeguards.
Recommended defensive actions
- Ensure affected monitors are connected to the internet so the automatic security update process can run.
- Maintain possession of the home monitor and restrict physical access to the device.
- Use only home monitors provided directly by a healthcare provider or a Medtronic representative.
- Verify whether deployed devices are model 24950 or 24952 and confirm they have received the vendor update.
- Follow Medtronic's security bulletin and contact [email protected] if additional assistance is needed.
- Apply CISA guidance for securing internet-connected devices and home network security in patient home deployments.
Evidence notes
All substantive claims here come from the supplied CISA CSAF advisory and its remediation text. The advisory for ICSMA-25-205-01 / CVE-2025-4394 was initially published on 2025-07-24 and revised in Update A on 2026-05-07. CISA states the vulnerability involves an unencrypted filesystem on internal storage and that exploitation requires physical tampering. The remediation section says Medtronic began deploying security updates in June 2025 and that the update process runs automatically when the monitor is connected to the internet.
Sources and references
Verified primary and authoritative sources
-
CVE-2025-4394 CVE Program record
Publisher, destination, and source semantics verified
URL: https://www.cve.org/CVERecord?id=CVE-2025-4394
CVE Program - Official CVE Program record with source-provided CVE metadata.
-
CVE-2025-4394 NVD vulnerability detail
Publisher, destination, and source semantics verified
URL: https://nvd.nist.gov/vuln/detail/CVE-2025-4394
NIST National Vulnerability Database - Official NIST NVD detail page and source-specific vulnerability assessment.
Supplemental references
-
Source item URL
Unverified legacy reference
URL: https://raw.githubusercontent.com/cisagov/CSAF/develop/csaf_files/OT/white/2025/icsma-25-205-01.json
cisa_csaf
-
Source reference
Unverified legacy reference
URL: https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-205-01
Reference
-
Source reference
Unverified legacy reference
URL: https://www.cisa.gov/resources-tools/resources/ics-recommended-practices
Reference
-
Source reference
Unverified legacy reference
URL: https://www.cisa.gov/sites/default/files/publications/Cybersecurity_Best_Practices_for_Industrial_Control_Systems.pdf
Reference
-
Source reference
Unverified legacy reference
URL: https://www.cisa.gov/topics/industrial-control-systems
Reference
-
Source reference
Unverified legacy reference
URL: https://www.cisa.gov/news-events/ics-alerts/ics-alert-10-301-01
Reference
-
Source reference
Unverified legacy reference
URL: https://www.cisa.gov/sites/default/files/recommended_practices/NCCIC_ICS-CERT_Defense_in_Depth_2016_S508C.pdf
Reference
-
Source reference
Unverified legacy reference
URL: https://www.cisa.gov/news-events/news/targeted-cyber-intrusion-detection-and-mitigation-strategies-update-b
Reference
Methodology and review provenance
AI-assisted synthesis based on stored public vulnerability evidence. System validation, approval state, and publication status do not by themselves establish human review of this revision. PatchSiren helps prioritize defensive review and does not prove exposure or remediation on any system.