PatchSiren cyber security CVE debrief
CVE-2025-4393 Medtronic CVE debrief
CVE-2025-4393 affects Medtronic MyCareLink Patient Monitor model 24950 and 24952. CISA says an internal service deserializes data, and a local attacker who can physically tamper with the monitor may craft a binary payload that can crash the service or elevate privileges. Medtronic says it began deploying security updates in June 2025, with automatic update delivery when the monitor is connected to the internet.
- Vendor
- Medtronic
- Product
- MyCareLink Patient Monitor model 24950
- CVSS
- MEDIUM 6.5
- CISA KEV
- Not listed in stored evidence
- Original CVE published
- 2025-07-24
- Original CVE updated
- 2026-05-07
- Advisory published
- 2025-07-24
- Advisory updated
- 2026-05-07
Who should care
Patients using the affected MyCareLink home monitor, clinicians who prescribe or support it, and biomedical/IT staff responsible for connected home-monitor deployments should pay attention, especially for model 24950 and model 24952.
Technical summary
The supplied CISA CSAF advisory describes an internal service that deserializes data. The stated attack path requires local access/physical tampering with the device, after which a crafted binary payload could interact with the service to cause a crash or privilege escalation. The advisory references CWE-502 and notes Medtronic began deploying security updates in June 2025.
Defensive priority
Medium. The impact includes service disruption and possible privilege escalation, but the advisory also says exploitation would require physical tampering/local access and Medtronic describes the findings as low-risk.
Recommended defensive actions
- Ensure the monitor is connected to the internet so the automatic security update process can complete.
- Keep possession of the home monitor and do not leave it accessible to untrusted parties.
- Use only home monitors provided directly by a healthcare provider or a Medtronic representative.
- If you need assistance, contact Medtronic at [email protected].
- Follow CISA guidance for securing internet-connected devices and home networks.
- For clinical workflows, continue prescribing and using the monitor as intended while ensuring affected units receive updates.
Evidence notes
All core claims come from the supplied CISA CSAF advisory and its embedded remediation text. The advisory revision history shows initial publication on 2025-07-24 and Update A on 2026-05-07. The remediation text states that security updates are applied automatically when the monitor is connected to the internet, and that users should maintain possession of the home monitor and only use monitors provided directly by a healthcare provider or Medtronic representative.
Official resources
-
CVE-2025-4393 CVE record
CVE.org
-
CVE-2025-4393 NVD detail
NVD
-
Source item URL
cisa_csaf
-
Source reference
Reference
-
Source reference
Reference
-
Source reference
Reference
-
Source reference
Reference
-
Source reference
Reference
-
Source reference
Reference
-
Source reference
Reference
CVE-2025-4393 was published on 2025-07-24, and the advisory was later revised as Update A on 2026-05-07. The supplied data does not indicate a CISA KEV listing or known ransomware campaign use.